Sending and Receiving Clock Settings
M3860A and M3861A H
EARTSTREAM
FR2 Defibrillator
6-9
6
Sending and Receiving Clock Settings
To synchronize the clock settings of your H
EARTSTREAM
FR2 with the clock of
another FR2 or a computer running C
ODE
R
UNNER
software, you can use the infrared
communication feature.
Instructions for synchronizing clock settings using a computer running C
ODE
R
UNNER
are provided with the C
ODE
R
UNNER
software.
To transfer clock settings from one FR2 to another:
1.
Remove and reinstall the battery of both FR2 devices to bring up the first menu
screen.
2.
Select NEXT to go to the second menu screen.
3.
Select CLOCK from the second menu screen. The CLOCK screen then comes up.
4.
Locate the infrared communication port on each H
EARTSTREAM
FR2 and line
them up with one another, so that the infrared “eye” in each one has an
uninterrupted view of the “eye” in the other. (See the diagram on the back of the
first page of this manual.) The two devices should be no more than 1 meter
apart.
5.
Select SEND TIME from the CLOCK screen on the “sending” H
EARTSTREAM
FR2.
6.
Select RECEIVE TIME from the CLOCK screen of the “receiving” FR2.
7.
A new screen comes up. Until the two H
EARTSTREAM
FR2 devices are properly
positioned, the screen on the receiving FR2 displays READY TO RECEIVE and
prompts you to check the sending FR2. The screen on the sending FR2 displays
READY TO SEND and prompts you to check the receiving FR2.
8.
Clock settings are automatically transferred as soon as the infrared ports are
correctly aligned.
Using Advanced Mode Features
The H
EARTSTREAM
FR2 provides an advanced mode that allows responders who are
appropriately trained to override the programmed FR2 protocol and take
responsibility for certain aspects of the operating sequence used by the FR2 to treat
the patient.
As described earlier in this chapter, the factory default setup of the FR2 must be
modified to provide access to advanced mode features. This requires use of the
administration function of the M3864A Training & Administration Pack.
If you are an expert user authorized by your Medical Director to modify setup, hold
down both the Option buttons while installing the Training & Administration Pack
RUN SELFTEST
REVIEW INCIDENT
NO DATA CARD
GOOD BATTERY
NEXT
IN EMERGENCY
PRESS OFF TO QUIT
CLOCK
RETURN
RECEIVE TIME
SEND TIME
07 / 09 / 00
DD / MM / YY
14 : 28
RECEIVE TIME
EXIT
READY TO RECEIVE
CHECK SENDER
Содержание Heartstream FR2
Страница 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Страница 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Страница 9: ...iv Contents Notes...
Страница 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Страница 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Страница 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Страница 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Страница 65: ...Accessories A 2 Notes...
Страница 77: ...Symbols and Controls D 4 Notes...