Agilent Technologies Helium Leak Detector
DRAFT 12/1
1
/17
99
NOTE
The smallest mass flow meter that may be used while still
functioning. It could be as small as 300 SCCM, though this flow
rate may not be capable of reading mass flow when measuring
flows of mobile unit leak detectors (units with larger roughing
pumps).
Alternatively, if using a mass flow meter that is not supported by the leak detector software
(anything other than Alicat Scientific brand mass flow meter):
1. Attach the mass flow meter in series above the Agilent power probe sniffing gun (Agilent
part number K9565306).
2. Attach the mass flow meter to the sniffing gun through use of a KF to compression port
adapter, 1/8" to NW25 (Agilent part number KCP250125), if using the power probe as
the metering valve (Figure 5-22).
3. Perform an internal calibration.
4. Perform an external calibration using the sniffing probe by:
a. Selecting
Calibration > Calibrate With > External Leak.
b. Entering the value of the external calibrated leak. Keep the probe at the inlet of the
standard leak.
c. Pressing
Calibrate.
Figure 5-22 External Sniff Calibration Setup
5. Select
Test Configurations > Applications Setup
.
6. Select
PPM Sniff Method > Application Setup.
7. Ensure the system is in
Test
and the test port pressure adjusted between 5E-1 Torr 1 Torr.
8. Select the
MFM
option.
9. Input the flow rate - SCCM displayed from the MFM/MFC controller.
The leak detector is now ready to measure the concentration of Helium in PPM.
To make the connection between the power probe sniffing gun and the mass flow meter:
1. Fit the mass flow meter on the low-pressure end with an NW25 vacuum fitting
(Figure 5-23).
Содержание G8610A
Страница 4: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 10: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 12: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 24: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 54: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 64: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 102: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 170: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 174: ...Agilent Technologies Helium Leak Detector This page intentionally left blank ...
Страница 175: ...Agilent Technologies Helium Leak Detector This page intentionally left blank ...