Agilent Technologies Helium Leak Detector
22
DRAFT 12
/1
1/17
❑
Allow an adequate distance (2 to 3’) from the back of the
leak detector to the power source. This ensures easy access
to the power cord for connecting or disconnecting purposes.
Spectrometer
WARNING
Store the Ion Source/Preamplifier sub-assembly in a cool, dry
area in a tightly sealed, ESD protected container. Wear lint-free
gloves when handling the spectrometer. Wash hands
thoroughly after handling the spectrometer filaments and
especially before smoking or eating.
CAUTION
The spectrometer operates at a very high vacuum produced by
the high vacuum pump. Service of the spectrometer requires
that this vacuum be vented to the atmosphere.
CAUTION
Do not use grease or any other substance on O-rings or metal
seals that will come in contact with the spectrometer.
CAUTION
If the spectrometer magnet comes in contact with a magnetic
surface, the magnet can become degraded causing the
spectrometer to lose sensitivity.
Pumps
WARNING
To avoid injury, use proper lifting techniques when moving
pumps. Two people may be required to move pumps safely.
WARNING
The vacuum pumps are also compressors; incorrect operation
may be dangerous. Study the Mechanical Pump Operation
Manual enclosed with your pump before starting pumps.
The pumps are designed to prevent any thermal risk for user
safety. However, specific operating conditions may generate
temperatures >70
C.
Hot oil burns the skin. Service of the pumps in this area must be
performed by authorized personnel only. Stand back from a
mechanical pump before starting it.
Содержание G8610A
Страница 4: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 10: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 12: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 24: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 54: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 64: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 102: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 170: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 174: ...Agilent Technologies Helium Leak Detector This page intentionally left blank ...
Страница 175: ...Agilent Technologies Helium Leak Detector This page intentionally left blank ...