Agilent Technologies Helium Leak Detector
DRAFT 12/1
1
/17
129
a
7.4.1.5 Current Setup
The current setup menu displays all activated parameters including selected applications.
7.4.1.6 Turn Off
Figure 7-16 Power Off
This feature provides a soft shut down for the CPU and turbo pump prior to turning off the
main power/circuit breaker switch.
NOTE
This power off feature provides a two-minute braking
procedure for the turbo pump. This aides in turbo pump life
and for safe transportation of the leak detector
Range Stop
See Section 4 “Calibration Screen”
on page 65.
Range Stop
-
[Exp]
System is set to display the most
sensitive leak rate range, 10
-
[Exp]
.
The range stop line shows only if
range stop is on.
Reject Set Point
See Section 7.4.1 “Current Setup
Menu” on page 128.
Reject# 1,2,3,4 Active Reject set points displayed are active;
this line is blank when all set points
are disabled or inactive.
Split Flow
See Section 5.1.3.9 “Split Flow” on
page 89.
Shows split flow, only if the feature is
turned on. To change the status, go to
the Advanced Parameters screen.
Vent Lockout Status
See Section 3.1.2 “Menus Screen” on
page 63.
Shows the status of the vent lockout
feature only if it is on.
To change the
status of the vent lockout feature, go
to the Control Panel access screen.
Table 7-3 Current Setup Screen Conditions (Continued)
Condition
Display
Description
Содержание G8610A
Страница 4: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 10: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 12: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 24: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 54: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 64: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 102: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 170: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 174: ...Agilent Technologies Helium Leak Detector This page intentionally left blank ...
Страница 175: ...Agilent Technologies Helium Leak Detector This page intentionally left blank ...