Agilent Technologies Helium Leak Detector
154
DRAFT 12
/1
1/17
11
REJECT4_
OUT
O
Level
Active HIGH when the test
port pressure is over set
point #4.
Indicates that the value of
the test port pressure
exceeds the value set in the
4REJECT variable.
12
CAL_OUT_
OK
O
Level
Active HIGH when
calibration is successful.
Indicates the calibration has
successfully completed. Set
low on power-up or if
calibration is not
successfully completed.
13
AUTO_
MANUAL_
RANGE_IN
I
Level
Active HIGH when Manual
Ranging is selected.
Activation causes the leak
detector to change from
AUTO to MANUAL ranging
; Default is AUTO.
14
CAL_IN
I
> 200
ms Pulse
Active HIGH pulse.
Initiate CALIBRATING
routine with Internal
Calibrated Leak; Zeroing
routine is included.
16
ZERO_IN
I
> 200
ms Pulse
Active HIGH pulse.
Initiate ZEROING routine.
18
HOLD_IN
I
> 200
ms Pulse
Active HIGH pulse.
Initiate HOLD routine and
isolate the test port and any
test objects from leak
detector vacuum system.
20
VENT_IN
I
> 200
ms Pulse
Active HIGH pulse.
Initiate VENT routine and
vent the test port to
atmosphere.
22
TEST_IN
I
> 200
ms Pulse
Active HIGH pulse.
Initiate ROUGHING routine
on the test port and any
attached test objects.
Table A-6 I/O Connector Pin Signals (Continued)
Pin
No.
Signal Name
I/O
Signal
Description
Leak Detector Behavior
Содержание G8610A
Страница 4: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 10: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 12: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 24: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 54: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 64: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 102: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 170: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 174: ...Agilent Technologies Helium Leak Detector This page intentionally left blank ...
Страница 175: ...Agilent Technologies Helium Leak Detector This page intentionally left blank ...