Agilent Technologies Helium Leak Detector
132
DRAFT 12
/1
1/17
Cleanliness is vital when servicing the leak detector or any vacuum equipment. There are
some techniques more important in leak detector servicing than in general vacuum work:
CAUTION
Do not use silicone
oil or silicone grease.
Use powder-free butyl or polycarbonate gloves to prevent skin
oils from getting on vacuum surfaces.
Do not clean any aluminum parts with Alconox
®
. Alconox is
not compatible with aluminum and will cause damage.
NOTE
Agilent does not recommend the use of vacuum grease.
Vacuum grease absorbs helium tracer gas and releases this
slowly resulting in helium background during leak test
operations. If it must be used, use it sparingly and avoid
silicone types. Apiezon
®
L grease is recommended (Agilent Part
No. 695400004).
When removing, checking or replacing O-rings:
CAUTION
Remove O-rings carefully with your fingers. Do not use metal
tools for this task. This prevents scratching of any sealing
surfaces.
Wipe all O-rings clean with a lint-free cloth before installation
to ensure that no foreign matter is present to impair the seal.
Do not use grease or any other substances on O-rings or metal
seals that will come in contact with the spectrometer.
Do not use alcohol, methanol or other solvents on O-rings as
this causes deterioration and reduces their ability to hold a
vacuum.
NOTE
If applicable, apply a small amount of Apiezon
L grease and
wipe the O-rings shiny dry.
NOTE
Due to the effective cleaning nature of VacuSolv solvent and
its residue-free properties, Agilent’ Component and
Spectrometer Cleaning Kit (Part Number 670029096), used in
accordance with the kit instructions, is recommended for
cleaning spectrometer components. The kit can also be used for
fine cleaning of other parts in the leak detector’s vacuum
system such as valves and fittings. No rinsing steps or
high-temperature drying are required following cleaning with
VacuSolv. Although appropriate precautions are advised,
VacuSolv is compatible with most materials and does not
contain toxic chemicals or CFCs (chlorofluorocarbons). Other
acceptable solvents are isopropyl alcohol (IPA) or Dow
Corning
®
OS-20.
Содержание G8610A
Страница 4: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 10: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 12: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 24: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 54: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 64: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 102: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 170: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 174: ...Agilent Technologies Helium Leak Detector This page intentionally left blank ...
Страница 175: ...Agilent Technologies Helium Leak Detector This page intentionally left blank ...