Agilent Technologies Helium Leak Detector
148
DRAFT 12
/1
1/17
INIT-4REJECT
Sets the test port pressure value for Reject Set Point #4. Preceded by a
two-digit leak rate mantissa with a decimal point after the first digit
followed by
E-
, and then by a two-digit exponent: the test port TC pressure
in millitorr.
Example 1.5E-02 INIT-4REJECT
INIT-AREJECT
Same as INIT-1REJECT, but for the Audio Reject Set Point.
HIGH-xREJECT
Sets the reject set point to activate when the actual test value is higher than
the set point. Where x is a number from 1 to 4 or the letter A. For example,
the reject set point is set for 7.0E-05 and the actual test value is 9.2E-05.
LOW-xREJECT
Sets the reject set point to activate when the actual test value is lower than
the set point. Where x is a number from 1 to 4 or the letter A. For example,
the reject set point is set for 7.0E-05 and the actual test value is 4.3E-07.
INIT-ROUGH
Sets the Auto Sequencer rough time in seconds. Preceded by
XX
.
INIT-SYSLEAK
Sets the split flow leak value, system must be in TEST and the split flow
feature must be OFF when this parameter is set. Preceded by a two-digit
leak rate mantissa with a decimal point after the first digit followed by
E-
,
and then by a two-digit leak range exponent: the helium leak rate number
in atm cc/sec.
Example 3.4E-08 INIT-SYSLEAK.
INIT-STDLEAK
Sets the value of the internal leak. Preceded by a two-digit leak rate
mantissa with a decimal point after the first digit followed by
E-
, and then
by a two-digit leak range exponent: the helium leak rate number in atm
cc/sec.
Example: 3.2E-08 INIT-STDLEAK.
INIT-TCTPZERO
Used to calibrate the low-pressure end of the test port thermocouple to a
known pressure higher than zero. Preceded by a number (
XX
) of mTorr
representing the test port pressure at the time of the command.
INIT-TEMPFACTOR
Sets the temperature correction factor that is printed on the internal cali-
brated leak label. Preceded by –10 to +10.
Example: -7 INIT-TEMPFACTOR
INIT-TEST
Sets the Auto Sequencer test time in seconds. Preceded by
XX
.
LOCK-MS
Enables the MS lock feature. If the system exceeds 150 mTorr, the system
goes from Fine test directly into Roughing mode (no Contra mode). If the
leak rate exceeds the 7 range, the over range symbol appears.
Table A-2 Non-Volatile Operating Parameters (Continued)
Command
Parameter
Содержание G8610A
Страница 4: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 10: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 12: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 24: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 54: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 64: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 102: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 170: ...Agilent Technologies Helium Leak Detector This page intentionally left blank D R A F T 1 2 1 1 1 7 ...
Страница 174: ...Agilent Technologies Helium Leak Detector This page intentionally left blank ...
Страница 175: ...Agilent Technologies Helium Leak Detector This page intentionally left blank ...