The Keypad
User Information
53
Enter a setpoint value of 0 to turn the zone off. The Headspace
Sampler displays
OFF
.
As a general rule, the temperatures for the loop and transfer
line should be set higher than the oven temperature to avoid
condensation. Set the loop temperature 10 to 15 °C higher than
the oven, and set the transfer line 20 to 25 °C higher than the
oven.
Table 3
Zone temperature ranges
Heated zone
Allowed setpoint range
Default value (°C)
Oven
40–230 in 1 °C increments *
50
Loop
45–250 in 1 °C increments
60
Tr. Line
50–250 in 1 °C increments
70
*The minimum oven temperature is 40 °C or 10 °C above ambient, whichever is
greater.
C A U T I O N
Oven temperature should not exceed 10 °C below the boiling point of
any solvents in the vial. Exceeding this may cause the cap and septum
to leak or to pop off ; the vial pressure line could also become
contaminated.
medium_standard.book Page 53 Tuesday, February 17, 2004 10:14 AM
Содержание G1888
Страница 12: ...12 User Information medium_standard book Page 12 Tuesday February 17 2004 10 14 AM...
Страница 24: ...24 User Information Introduction medium_standard book Page 24 Tuesday February 17 2004 10 14 AM...
Страница 118: ...118 User Information Developing Methods medium_standard book Page 118 Tuesday February 17 2004 10 14 AM...
Страница 162: ...162 User Information Maintaining the Instrument medium_standard book Page 162 Tuesday February 17 2004 10 14 AM...
Страница 190: ...190 User Information Index medium_standard book Page 190 Tuesday February 17 2004 10 14 AM...
Страница 191: ...medium_standard book Page 191 Tuesday February 17 2004 10 14 AM...