![Agilent Technologies G1888 Скачать руководство пользователя страница 135](http://html1.mh-extra.com/html/agilent-technologies/g1888/g1888_user-information_2867733135.webp)
Maintaining the Instrument
User Information
135
4
Exchange the sample loop.
a
Slide the loop off the mandrel.
b
Slide the new sample loop onto the
mandrel.
•
When installing a deactivated
sample loop, take care not to bend
the loop when installing it or you
will break the coating and
compromise inertness.
5
Reverse the steps given above to
reassemble the Headspace
Sampler.
•
Perform a leak test on fittings
before replacing the insulation and
manifold cover.
Table 12
Changing the sample loop (continued)
Step
Action
Notes
Port 1
Port 4
medium_standard.book Page 135 Tuesday, February 17, 2004 10:14 AM
Содержание G1888
Страница 12: ...12 User Information medium_standard book Page 12 Tuesday February 17 2004 10 14 AM...
Страница 24: ...24 User Information Introduction medium_standard book Page 24 Tuesday February 17 2004 10 14 AM...
Страница 118: ...118 User Information Developing Methods medium_standard book Page 118 Tuesday February 17 2004 10 14 AM...
Страница 162: ...162 User Information Maintaining the Instrument medium_standard book Page 162 Tuesday February 17 2004 10 14 AM...
Страница 190: ...190 User Information Index medium_standard book Page 190 Tuesday February 17 2004 10 14 AM...
Страница 191: ...medium_standard book Page 191 Tuesday February 17 2004 10 14 AM...