Maintaining the Instrument
User Information
157
Recommended tools
You will need the tools listed in
to complete the zone
calibration procedure.
Procedure
The procedure consists of three tasks:
1
Collect the temperature measurements
2
Enter the measurements into the Headspace Sampler
3
Store the custom offset
.
Table 21
Recommended tools
Qty
Description
Notes
1
20-mL sample vial and cap
Recommend vial part number 5182-0837 and cap and
septum part number 5183-4477.
1
No.1 Phillips head screw driver
Used for accessing the loop heated zone.
1
High temperature tape
Used to secure the probe lead from the oven.
1
Timer
Used for tracking the equilibration time between the zone
reaching the set point and the measurement.
1
Digital thermometer
Recommend 2 inputs with accuracy of ±0.2 °C.
2
Type K thermocouple bead probe
Able to function between 0 °C and 260 °C with an
accuracy of ±1.1 °C
medium_standard.book Page 157 Tuesday, February 17, 2004 10:14 AM
Содержание G1888
Страница 12: ...12 User Information medium_standard book Page 12 Tuesday February 17 2004 10 14 AM...
Страница 24: ...24 User Information Introduction medium_standard book Page 24 Tuesday February 17 2004 10 14 AM...
Страница 118: ...118 User Information Developing Methods medium_standard book Page 118 Tuesday February 17 2004 10 14 AM...
Страница 162: ...162 User Information Maintaining the Instrument medium_standard book Page 162 Tuesday February 17 2004 10 14 AM...
Страница 190: ...190 User Information Index medium_standard book Page 190 Tuesday February 17 2004 10 14 AM...
Страница 191: ...medium_standard book Page 191 Tuesday February 17 2004 10 14 AM...