Maintaining the Instrument
User Information
153
8
Allow 5 minutes to pass. Record the final time and pressure
values in
. Carefully close the lid of the sampler so
the sample probe enters the leak test vial.
A pressure decay of more than 2 psi over 5 minutes indicates a
leak. If there is a leak, call an Agilent service representative. Fix
any leaks before continuing the leak test.
Pressure decay test part 3: Inject mode
This pressure decay test checks for leaks in inject mode.
Perform the following steps to run the pressure decay test:
1
Open the shutoff valves on the back of the Headspace
Sampler.
2
Wait 5 minutes for the flow paths to pressurize and
equilibrate.
3
Turn off the shutoff valves on the back of the Headspace
Sampler.
4
. Record the time and initial pressure values in
the table.
Table 16
Pressure decay test part 2B
Final time
Initial time
Difference
Time
5 minutes
Flow path
Initial pressure
Final pressure
Carrier
Table 17
Pressure decay test part 3
Final time
Initial time
Difference
Time
5 minutes
Flow path
Initial pressure
Final pressure
Carrier
Vial pressurization
medium_standard.book Page 153 Tuesday, February 17, 2004 10:14 AM
Содержание G1888
Страница 12: ...12 User Information medium_standard book Page 12 Tuesday February 17 2004 10 14 AM...
Страница 24: ...24 User Information Introduction medium_standard book Page 24 Tuesday February 17 2004 10 14 AM...
Страница 118: ...118 User Information Developing Methods medium_standard book Page 118 Tuesday February 17 2004 10 14 AM...
Страница 162: ...162 User Information Maintaining the Instrument medium_standard book Page 162 Tuesday February 17 2004 10 14 AM...
Страница 190: ...190 User Information Index medium_standard book Page 190 Tuesday February 17 2004 10 14 AM...
Страница 191: ...medium_standard book Page 191 Tuesday February 17 2004 10 14 AM...