Running a Sample
User Information
123
Setting the vial pressurization gas pressure using MPC
Control the gas pressure used for vial pressurization using the
pressure regulator. See
Set the vial pressurization gas higher than the pressure inside
of the vial to prevent a backwards flow from contaminating the
vial pressurization flow path. A setting of 10 to 20 psi is usually
high enough.
To read the actual pressure value, press
Menu
and choose
Pressure
from the menu. The pressure displays in the right
side of the window. See
.
Setting carrier flow using EPC
Set the carrier gas flow rate directly using the GC or GC data
system. An aux EPC module is required when using the 6890
GC.
Figure 18
Measuring the split vent flow rate
Electronic flow meter
Split vent
6850 GC shown
medium_standard.book Page 123 Tuesday, February 17, 2004 10:14 AM
Содержание G1888
Страница 12: ...12 User Information medium_standard book Page 12 Tuesday February 17 2004 10 14 AM...
Страница 24: ...24 User Information Introduction medium_standard book Page 24 Tuesday February 17 2004 10 14 AM...
Страница 118: ...118 User Information Developing Methods medium_standard book Page 118 Tuesday February 17 2004 10 14 AM...
Страница 162: ...162 User Information Maintaining the Instrument medium_standard book Page 162 Tuesday February 17 2004 10 14 AM...
Страница 190: ...190 User Information Index medium_standard book Page 190 Tuesday February 17 2004 10 14 AM...
Страница 191: ...medium_standard book Page 191 Tuesday February 17 2004 10 14 AM...