Start/Stop key
Press this key to start or stop a run. To start a run, press once.
To stop a run, press the key again. The display reads:
PRESS
STOP
TO CONFIRM
This prevents an inadvertent run interruption. To stop, press
the key again. To continue the run, press
Clear
.
•
When the sampler stops during a run, the vials in the oven
automatically return to the sample tray. The method cannot
be resumed. To continue the analysis from the last vial
sampled, change the method so that it starts with the correct
vial.