Maintaining the Instrument
User Information
137
3
Remove the two screws.
•
Remove the two Phillips head screws
shown below.
4
Disconnect the plumbing
a
Disconnect the solenoid valves. Use
an 8-mm wrench to secure the
connection at the solenoid fitting and a
7-mm wrench to turn the connection
on the tubing.
b
Disconnect the fitting from port 6 of
the 6-port valve. Use a 7-mm wrench.
Table 13
Replacing the tubing assembly (continued)
Step
Action
Notes
Screws
Connections to
solenoid valves
Connection to
6-port valve
medium_standard.book Page 137 Tuesday, February 17, 2004 10:14 AM