49
3 Installing the BenchBot Robot workstation
Installing and removing the robot
BenchBot Robot Safety and Installation Guide
2
Turn off all devices in the workstation. For instructions, see the device
user documentation.
3
Turn off the robot. See the
BenchBot Robot User Guide
.
Procedure
To disconnect the robot:
Disconnect and remove the following:
•
Power cable
•
Ethernet cable
•
Robot- hub cable (G5486A BenchBot Robot), or pendant cable (G5487A
BenchBot Robot)
To detach the robot:
1
Using the supplied 5- mm hex wrench, unscrew the four M6 screws that are
holding the robot to the integration plate or attachment surface.
2
Place the robot on a flat, stable surface.
Related information