3 Installing the BenchBot Robot workstation
Assembling and disassembling the integration plates
BenchBot Robot Safety and Installation Guide
b
Make sure that the counter- sink holes are aligned over the threaded
holes on the lips of adjacent plates.
c
Position the plate assembly so that the clearance holes in the plates
align over the mounting holes in the attachment surface.
2
Lock the plates together. Insert the FHSC M5 x 12 screws in the counter-
sink holes, and use the 4- mm hex wrench to tighten them until snug.
3
Secure the assembled plates to the attachment surface. See your facilities
personnel for instructions.
4
Peel the protective paper from the back of the drip sheets, and then
carefully place the sheets on the integration plates. The drip sheets prevent
spills from leaking below the integration plates.
5
Mark the placement of robots and devices on the plate assembly. For
example, you can use removable tape to mark the positions before
mounting the robots and devices.
To attach the BenchBot Robot, see
“Installing and removing the robot” on
. To attach other devices, see the device user guides.
Disassembling the integration plates
To disassemble the integration plates:
1
Remove the drip sheets from the integration plates.
2
Using the 4- mm hex wrench, loosen and remove the FHSC M5 x 12 screws
in the counter- sink holes.
3
Consult your facilities personnel for instructions on removing the
integration plates from the attachment surface.
4
Store the plates in the original shipping box.
5
Place the screws in a bag and store them with the plates.
Содержание BenchBot
Страница 1: ...BenchBot Robot Safety and Installation Guide Original Instructions...
Страница 4: ...Contents iv BenchBot Robot Safety and Installation Guide...
Страница 10: ...Preface Reporting problems x BenchBot Robot Safety and Installation Guide...
Страница 71: ......
Страница 72: ...Safety and Installation Guide G5486 90001 Revision 01 June 2011...