2 Laboratory setup requirements
BenchBot integration plate specifications
20
BenchBot Robot Safety and Installation Guide
Figure
The mounting surface and the lip
Plate assembly requirements
M5 x 12 flat- head socket- cap screws (FHSC screws, supplied) are required to
lock adjacent plates together. Use only the counter- sink holes as shown in the
diagram. For assembly instructions, see
“Installing the robot” on page 27
.
Figure
Countersink holes for joining adjacent plates
Plate mounting specifications
M5 FHSC screws (supplied) are required to secure the integration plates to the
table surface. Use only the clearance holes on the edges for securing the plate
assembly to the table surface.
1
2
Mounting requirement
Measurement
Hole distance (center to center)
76 mm (3 in)
Screw type
M5 FHSC screw*
* Length and type depends on the table
surface material.