Agilent E5250A User’s Guide, Edition 11
9-19
Executing Sample Programs
HCI Measurement Program
Step 3. Selects devices valid for the HCI test
This step judges whether each device is valid according to the current leak test
result. The
IGLEAK.MES
and
IDLEAK.MES
files are used to set up the 4155/4156
for this step.
As shown below, 24 devices are connected to the E5250A, and measured
individually by switching. For devices that fail this step, no further testing is
performed, and "0.0" is recorded.
Step 4. Characterizes the initial parameters
This step initially (before stress) measures parameters for the devices that were
determined to be valid by the previous step.
The
PARAM.MES
file sets up the 4155/4156.
A summary of the measurement setup saved in
PARAM.MES
is shown below. For
details of the setup data, GET the
PARAM.MES
file using the 4155/4156, and refer
to the setup displayed on the 4155/4156.
Channel/Source Setup:
Device
Terminal
Device 1
Device2
Device 3
. . . .
Device 24
Gate
1
2
3
. . . .
24
Drain
25
26
27
. . . .
48
Sub
49
50
51
. . . .
72
Source
GNDU
GNDU
GNDU
GNDU
GNDU
SMU
V NAME
I NAME
MODE
FUNCTION
Source Setup
SMU1
VG
IG
V
VAR1
0 to 5V, 20mV step
SMU3
VD
ID
V
CONST
0.1V
SMU4
VB
IB
V
CONST
0V
GNDU
VS
IS
COMMON
CONST
Common
Содержание Agilent E5250A
Страница 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Страница 10: ......
Страница 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Страница 17: ...1 Introduction ...
Страница 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Страница 33: ...2 Installation ...
Страница 51: ...3 Executing Self Test and Leak Test ...
Страница 70: ...3 20 Agilent E5250A User s Guide Edition 11 Executing Self Test and Leak Test Using the Self Test Utility ...
Страница 71: ...4 Setting up Measurement Environment ...
Страница 95: ...5 Controlling the E5250A ...
Страница 137: ...6 Programming the E5250A ...
Страница 169: ...7 Command Reference ...
Страница 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Страница 225: ...8 VXIplug play Driver ...
Страница 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Страница 257: ...9 Executing Sample Programs ...
Страница 295: ...10 Specifications ...
Страница 304: ...10 10 Agilent E5250A User s Guide Edition 11 Specifications Supplemental Data when used with Agilent 4155 4156 ...
Страница 305: ...11 Error Messages ...
Страница 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Страница 319: ...12 SCPI Command Summary ...