3-6
Agilent E5250A User’s Guide, Edition 11
Executing Self-Test and Leak Test
Executing Self-Test
To Execute Self-Test using External Controller
You can execute Self-Test from an external controller by using SCPI commands in a
program.
This section describes the requirements, SCPI commands, and a sample program for
Self-Test execution from an external controller. For details about the SCPI
commands, refer to Chapter 7.
Requirements
•
Controller
•
GPIB Cable
•
Relay Test Adapter
•
BNC Short Caps (3 caps are furnished with each E5255A)
SCPI Commands for Self-Test
Table 3-2 shows the SCPI commands used to execute Self-Test. Keyword of the
SCPI command first level is :DIAG.
The :DIAG commands ignore the E5250A channel configuration mode (NORMAL
or AUTO).
For :DIAG commands that require a card number, you specify
1
,
2
,
3
,
4
, or
ALL
.
If the Fail LED turns on after executing the Self-Test, the E5250A or modules may
be defective. Contact your nearest Agilent Technologies Service Center.
NOTE
Before executing Relay Test, be careful about following:
• Attach the Relay Test Adapter to the E5250A input connectors. Refer to Figure
3-1. Then set the switch on the adapter to the
Relay Test
position.
• Open all outputs of the plug-in cards as follows: If cable is connected to output
connector of plug-in card, open output at DUT end of cable. If end of cable is
Kelvin connection, disconnect the cable or make non-Kelvin connection at end
of cable.
• For each E5255A installed in the E5250A, connect a BNC short cap to each
E5255A BIAS INPUT connector.
Содержание Agilent E5250A
Страница 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Страница 10: ......
Страница 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Страница 17: ...1 Introduction ...
Страница 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Страница 33: ...2 Installation ...
Страница 51: ...3 Executing Self Test and Leak Test ...
Страница 70: ...3 20 Agilent E5250A User s Guide Edition 11 Executing Self Test and Leak Test Using the Self Test Utility ...
Страница 71: ...4 Setting up Measurement Environment ...
Страница 95: ...5 Controlling the E5250A ...
Страница 137: ...6 Programming the E5250A ...
Страница 169: ...7 Command Reference ...
Страница 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Страница 225: ...8 VXIplug play Driver ...
Страница 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Страница 257: ...9 Executing Sample Programs ...
Страница 295: ...10 Specifications ...
Страница 304: ...10 10 Agilent E5250A User s Guide Edition 11 Specifications Supplemental Data when used with Agilent 4155 4156 ...
Страница 305: ...11 Error Messages ...
Страница 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Страница 319: ...12 SCPI Command Summary ...