Agilent E5250A User’s Guide, Edition 11
8-29
VXIplug&play Driver
Driver Functions
hpe5250a_selectCompenFile
This function specifies capacitance/conductance compensation data file used to
compensate C/G by using hpe5250a_compenC.
Syntax
ViStatus _VI_FUNC hpe5250a_selectCompenFile(ViSession vi,
ViString file_name);
Parameters
vi
Instrument handle returned from hpe5250a_init( ).
file_name
Compensation data file name. Use absolute path. If the value is
NULL string, the default data is used.
Remarks
If you change the compensation data, copy the default data shown below, and
modify the data for your measurement cable. You will need to change the data for
DATA05 and 06, and/or DATA07 and 08 corresponding to your cables. To measure
and change the compensation data, refer to Chapter 6. To get the R, L, and C value,
measure R, L, and C of the cable using the 4284A, and divide them by cable length
(in m). Compensation data must be the value for 1 m length. Do not change the data
format in the file.
# E5250A C Compensation coefficient data table
#
# CAUTION : Do not add or delete "REVISION" line and "DATAxx" line.
# Change the value for R,L,C of DATA05,06,07 or 08.
#
REVISION
A.03.00
#
R [ohm]
L [H]
C [F]
DATA00
74.65E-3
140.00E-9
58.44E-12
# Frame Path 1
DATA01
75.41E-3
90.00E-9
67.13E-12
# Frame Path 2
DATA02
231.41E-3
450.00E-9
178.85E-12
# Card Path High
DATA03
177.56E-3
390.00E-9
135.45E-12
# Card Path Low
DATA04
100.70E-3
400.00E-9
80.00E-12
# Triax Cable [/m]
DATA05
100.70E-3
400.00E-9
80.00E-12
# User Triax Cbl H [/m]
DATA06
100.70E-3
400.00E-9
80.00E-12
# User Triax Cbl L [/m]
DATA07
114.00E-3
544.00E-9
130.00E-12
# User Coax Cbl H [/m]
DATA08
114.00E-3
544.00E-9
130.00E-12
# User Coax Cbl L [/m]
DATA09
0.00E-3
0.00E-9
1.20E-12
# Stray Capacitance
# END of Data
Содержание Agilent E5250A
Страница 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Страница 10: ......
Страница 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Страница 17: ...1 Introduction ...
Страница 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Страница 33: ...2 Installation ...
Страница 51: ...3 Executing Self Test and Leak Test ...
Страница 70: ...3 20 Agilent E5250A User s Guide Edition 11 Executing Self Test and Leak Test Using the Self Test Utility ...
Страница 71: ...4 Setting up Measurement Environment ...
Страница 95: ...5 Controlling the E5250A ...
Страница 137: ...6 Programming the E5250A ...
Страница 169: ...7 Command Reference ...
Страница 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Страница 225: ...8 VXIplug play Driver ...
Страница 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Страница 257: ...9 Executing Sample Programs ...
Страница 295: ...10 Specifications ...
Страница 304: ...10 10 Agilent E5250A User s Guide Edition 11 Specifications Supplemental Data when used with Agilent 4155 4156 ...
Страница 305: ...11 Error Messages ...
Страница 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Страница 319: ...12 SCPI Command Summary ...