Agilent E5250A User’s Guide, Edition 11
9-37
Executing Sample Programs
HCI Measurement Program
To change Idlin and Gmmax extraction condition
To change Gate voltage and Substrate voltage required for Idlin and Gmmax
extraction, do follows.
1. GET the PARAM.MES setup file by using the 4155/4156.
2. Change the value of Substrate voltage "VB" on the SOURCE SETUP screen.
3. Change the setting of "VG" parameter in "LINE1" on the ANALYSIS SETUP
screen.
To change the number of devices
Change the setting of the
No_of_devices
variable in the "Input Parameters" block.
To test more than 25 devices, you must change the configuration of the E5250A. See
chapter 2.
To change the stress times
Change the data at the DATA statement in the "Stress duration setup" block. Each
data shows the cumulative stress data.
Example:
DATA
10,
20,
50
DATA
100,
200,
500
DATA
1000,
2000,
5000
DATA
10000,
20000,
50000
DATA
100000
If you change the number of the data defined in DATA statements, change the value
of the
Meas_points
variable in the "Stress duration setup" block.
The value of
Meas_points
must be the same number as the number of data defined
in the DATA statements.
In the above example, the number of data in DATA statements is 13. So the variable
for the above example must be set as follows.
Meas_points=13
Содержание Agilent E5250A
Страница 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Страница 10: ......
Страница 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Страница 17: ...1 Introduction ...
Страница 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Страница 33: ...2 Installation ...
Страница 51: ...3 Executing Self Test and Leak Test ...
Страница 70: ...3 20 Agilent E5250A User s Guide Edition 11 Executing Self Test and Leak Test Using the Self Test Utility ...
Страница 71: ...4 Setting up Measurement Environment ...
Страница 95: ...5 Controlling the E5250A ...
Страница 137: ...6 Programming the E5250A ...
Страница 169: ...7 Command Reference ...
Страница 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Страница 225: ...8 VXIplug play Driver ...
Страница 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Страница 257: ...9 Executing Sample Programs ...
Страница 295: ...10 Specifications ...
Страница 304: ...10 10 Agilent E5250A User s Guide Edition 11 Specifications Supplemental Data when used with Agilent 4155 4156 ...
Страница 305: ...11 Error Messages ...
Страница 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Страница 319: ...12 SCPI Command Summary ...