Agilent E5250A User’s Guide, Edition 11
5-9
Controlling the E5250A
Switch Control Basics
Connection Rule
Connection rule is used to specify if an input or output port can have multiple
connections, and it is specified by the [:ROUTe]:CONNection:RULE command
described in Chapter 7. For each card installed in the E5250A (for auto
configuration mode, installed cards are treated as one card), you can specify one of
the following connection rules:
Single Route:
Each input port can be connected to
only one
output port, and
each output port can be connected to
only one
input port. So,
existing connection to a port must be disconnected when a new
connection is made.
If normal configuration mode is set to the E5250A with
multiple cards, an input/output path is effective for each card.
For example, if normal configuration mode is set to the E5250A
with two E5252As, and single route is set to both cards, the
following connection is possible.
Free:
Each input port can be connected to
multiple
output ports, and
each output port can be connected to
multiple
input ports.
CAUTION
If the Free connection rule has been specified, ensure multiple input ports are not
connected to the same output port. Such configurations can cause damage.
E5250A
INPUT
E5252A Card 1 Output
1
12
2
1
Close
Open
10
3
2
E5252A Card 2 Output
1
2
12
3
3
Содержание Agilent E5250A
Страница 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Страница 10: ......
Страница 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Страница 17: ...1 Introduction ...
Страница 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Страница 33: ...2 Installation ...
Страница 51: ...3 Executing Self Test and Leak Test ...
Страница 70: ...3 20 Agilent E5250A User s Guide Edition 11 Executing Self Test and Leak Test Using the Self Test Utility ...
Страница 71: ...4 Setting up Measurement Environment ...
Страница 95: ...5 Controlling the E5250A ...
Страница 137: ...6 Programming the E5250A ...
Страница 169: ...7 Command Reference ...
Страница 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Страница 225: ...8 VXIplug play Driver ...
Страница 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Страница 257: ...9 Executing Sample Programs ...
Страница 295: ...10 Specifications ...
Страница 304: ...10 10 Agilent E5250A User s Guide Edition 11 Specifications Supplemental Data when used with Agilent 4155 4156 ...
Страница 305: ...11 Error Messages ...
Страница 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Страница 319: ...12 SCPI Command Summary ...