Agilent E5250A User’s Guide, Edition 11
5-29
Controlling the E5250A
Using Virtual Front Panel Utility
Changing the VFP Setup Mode
Select the SETUP MODE softkey. The display shows the present setup conditions:
Channel Configuration mode:
NORMAL or AUTO
Model Number of each card:
E5252A or E5255A
Bias Mode state of each card:
ON or OFF
Bias Port number of each card:
E5252A: 1 to 10 E5255A: not applicable
The card numbers are 1 to 4 in NORMAL configuration mode, and 0 (zero) in
AUTO configuration mode.
The following softkeys are available.
CAUTION
If you change the VFP Setup Mode, all data is cleared from the setup memories
when you return to the VFP main menu.
Softkey Description
CHANGE
CONFIG
For changing the channel configuration mode. See next page.
CARDn
(n=1 to 4)
For selecting which card to change the Bias Port and Bias Mode
state. CARDn softkeys are displayed only in NORMAL
configuration mode. Displays the BIAS STATE, BIAS PORT and
RETURN softkeys.
Select RETURN softkey to display CARDn softkey again, so you
can select another card to change the Bias Port and Bias Mode state.
BIAS
STATE
For toggling the Bias Mode ON/OFF. This softkey is displayed in
AUTO configuration mode, or by selecting the CARDn softkey in
NORMAL configuration mode. See next page.
BIAS
PORT
Only for the E5252A card. For selecting the Bias Port. This softkey
is displayed in AUTO configuration mode, or by selecting the
CARDn softkey in NORMAL configuration mode. See next page.
MAIN
MENU
For returning to VFP main menu. See next page.
Содержание Agilent E5250A
Страница 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Страница 10: ......
Страница 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Страница 17: ...1 Introduction ...
Страница 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Страница 33: ...2 Installation ...
Страница 51: ...3 Executing Self Test and Leak Test ...
Страница 70: ...3 20 Agilent E5250A User s Guide Edition 11 Executing Self Test and Leak Test Using the Self Test Utility ...
Страница 71: ...4 Setting up Measurement Environment ...
Страница 95: ...5 Controlling the E5250A ...
Страница 137: ...6 Programming the E5250A ...
Страница 169: ...7 Command Reference ...
Страница 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Страница 225: ...8 VXIplug play Driver ...
Страница 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Страница 257: ...9 Executing Sample Programs ...
Страница 295: ...10 Specifications ...
Страница 304: ...10 10 Agilent E5250A User s Guide Edition 11 Specifications Supplemental Data when used with Agilent 4155 4156 ...
Страница 305: ...11 Error Messages ...
Страница 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Страница 319: ...12 SCPI Command Summary ...