Executing Sample Programs
HCI Measurement Program
Executing the Program
You should have already added the bias source control routine and set up the
measurement environment as described previously.
Before running the program, do following:
1. Set the GPIB address of instruments as follows:
•
4155/4156: 717
•
E5250A: 722
•
Bias Source: Set the GPIB addresses to match your control routine
(Address 701 to 703 are defaults used in
SP_MUX_M
program)
2. Turn on all instruments and the computer.
3. Boot up HP BASIC operating system.
4. Insert a diskette into the computer's disk drive. The diskette must contain the
HCI test program (SP_MUX in this example) you modified in “Adding the Bias
Source Control Routine” on page 9-23.
5. Specify the computer's mass storage device (disk drive) the diskette is inserted
by using MSI statement. For example, if the device is ":CS80,702,1", enter the
following command:
MSI ":CS80,702,1"
6. Load your HCI test program. The following example loads and displays
SP_MUX program, in this example.
LOAD "SP_MUX"
EDIT
7. Insert the E5250A Program Disk into the 4155/4156's disk drive, then get the
setup file
IDVD.MES
to set up the 4155/4156.
8. Enter the following command from the command line of the computer:
OUTPUT 722;":ROUT:CLOS (@10101,20101,30101)"
9. Press Single key on the 4155/4156's front panel. The Id-Vds measurement is
performed and the Id-Vds measurement result is displayed as shown in Figure
9-5.
10. Determine the drain stress voltage (
Vdstr
) from the above curve.
Recommended maximum value is about 0.5 V below actual breakdown. Then,
edit the HCI test program (SP_MUX in this example) , and define the
Vdstr
value. The value should be defined in the following program line.
Содержание Agilent E5250A
Страница 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Страница 10: ......
Страница 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Страница 17: ...1 Introduction ...
Страница 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Страница 33: ...2 Installation ...
Страница 51: ...3 Executing Self Test and Leak Test ...
Страница 70: ...3 20 Agilent E5250A User s Guide Edition 11 Executing Self Test and Leak Test Using the Self Test Utility ...
Страница 71: ...4 Setting up Measurement Environment ...
Страница 95: ...5 Controlling the E5250A ...
Страница 137: ...6 Programming the E5250A ...
Страница 169: ...7 Command Reference ...
Страница 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Страница 225: ...8 VXIplug play Driver ...
Страница 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Страница 257: ...9 Executing Sample Programs ...
Страница 295: ...10 Specifications ...
Страница 304: ...10 10 Agilent E5250A User s Guide Edition 11 Specifications Supplemental Data when used with Agilent 4155 4156 ...
Страница 305: ...11 Error Messages ...
Страница 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Страница 319: ...12 SCPI Command Summary ...