Agilent E5250A User’s Guide, Edition 11
9-23
Executing Sample Programs
HCI Measurement Program
Adding the Bias Source Control Routine
This sample program requires bias sources to apply stress to DUT. So, you must
modify the following bias source control subprograms to match your bias sources or
create new subprograms. The following subprogram is for bias sources that support
SCPI commands:
1. Get the HCI test program (SP_MUX_M) from the E5250A Program Disk by
using the computer:
GET "SP_MUX_M"
2. Delete the exclamation marks (!) from beginning of the following two program
lines. You can easily find these lines by using FIND statement.
! CALL Bias_source_on
:
:
! CALL Bias_source_off
3. Delete the exclamation mark (!) from the beginning of the following lines in the
"Bias_source_on" and "Bias_source_off" subprograms.
Then modify the "OUTPUT" statements to match your Bias Sources. The
"SOUR XXX" SCPI command is used to control the output voltage of the Bias
Sources.
!Bias_source_on:SUB Bias_source_on
! !===========================================================
! ! Bias Source On
! !-----------------------------------------------------------
! COM /Meas_info5/ REAL Vdstr,Vgstr,Vdd,Vbb,Gate_length,Gate_w
idth
! !-----------------------------------------------------------
! OUTPUT 701;"SOUR "&VAL$(Vgstr) ! Set Gate Stress Voltage
! OUTPUT 702;"SOUR "&VAL$(Vdstr) ! Set Drain Stress Voltage
! OUTPUT 703;"SOUR 0.0" ! Set Substrate Stress Voltage
!SUBEND
! !===========================================================
!Bias_source_off:SUB Bias_source_off
! !===========================================================
! ! Bias Source Off
! !-----------------------------------------------------------
! COM /Meas_info5/ REAL Vdstr,Vgstr,Vdd,Vbb,Gate_length,Gate_w
idth
! !-----------------------------------------------------------
! OUTPUT 701;"SOUR 0.0" ! Set Gate Voltage to 0.0V
! OUTPUT 702;"SOUR 0.0" ! Set Drain Voltage to 0.0V
! OUTPUT 703;"SOUR 0.0" ! Set Substrate Voltage to 0.0V
!SUBEND
4. Save modified file to an empty
formatted
diskette as in following example. This
diskette is used for executing program. And the data measured and analyzed will
be saved to this diskette automatically during program execution.
RE-STORE "SP_MUX"
Содержание Agilent E5250A
Страница 1: ...Agilent Technologies Agilent E5250A Low Leakage Switch Mainframe User s Guide ...
Страница 10: ......
Страница 16: ...Contents 6 Agilent E5250A User s Guide Edition 11 Contents 12 SCPI Command Summary ...
Страница 17: ...1 Introduction ...
Страница 32: ...1 16 Agilent E5250A User s Guide Edition 11 Introduction Options and Accessories ...
Страница 33: ...2 Installation ...
Страница 51: ...3 Executing Self Test and Leak Test ...
Страница 70: ...3 20 Agilent E5250A User s Guide Edition 11 Executing Self Test and Leak Test Using the Self Test Utility ...
Страница 71: ...4 Setting up Measurement Environment ...
Страница 95: ...5 Controlling the E5250A ...
Страница 137: ...6 Programming the E5250A ...
Страница 169: ...7 Command Reference ...
Страница 224: ...7 56 Agilent E5250A User s Guide Edition 11 Command Reference Output Queue ...
Страница 225: ...8 VXIplug play Driver ...
Страница 256: ...8 32 Agilent E5250A User s Guide Edition 11 VXIplug play Driver Driver Functions ...
Страница 257: ...9 Executing Sample Programs ...
Страница 295: ...10 Specifications ...
Страница 304: ...10 10 Agilent E5250A User s Guide Edition 11 Specifications Supplemental Data when used with Agilent 4155 4156 ...
Страница 305: ...11 Error Messages ...
Страница 318: ...11 14 Agilent E5250A User s Guide Edition 11 Error Messages E5250A Specific Error Messages ...
Страница 319: ...12 SCPI Command Summary ...