1260 Infinity Binary LC - System User Guide
91
Quick Start Guide
5
Data Acquisition in Method and Run Control View
Checkout Preparation
1
Fill solvent bottle A with HPLC grade Water. Fill solvent bottle B with
HPLC grade Acetonitrile.
2
If installed, purge each channel of the on- line degasser using the
supplied syringe.
3
Open the purge valve of the pump, and purge each channel at
5 mL/min.
This should be sufficient to remove air in the system.
4
Install the supplied column into the column compartment.
5
Flush the flow- cell with HPLC- grade water for 5 min.
6
Run a Wavelength Calibration and Intensity Test where applicable.
7
Allow the system to equilibrate at the checkout conditions listed in
on page 90 for 10 min.
8
Prepare a vial of the isocratic checkout sample.