1260 Infinity Binary LC - System User Guide
115
Appendix
6
Setting Up a Method using Edit Entire Method
Autosampler tab
Figure 25
Setup Method screen – High Performance Autosampler tab
• Injection Mode
•
Injection volume
sets the volume to be injected (example 1 µl),
•
Standard injection
indicates that no external needle wash is done,
•
Injection with needle wash
is used to reduce potential carryover. This is
the recommended option and is configured in the next entry.
• Needle Wash
– if selected above.
•
Mode
determines how the outside of the needle will be rinsed: either
actively in the
Flush Port
or by dipping into a specified
Wash Vial
.
•
Time
in seconds the peristaltic pump connected to the flush port will
pump the rinse solvent. It then pumps for a further 15 s to clear the
flush port.
•
Location
determines which vial or well plate will be used if
Wash Vial
was selected.
N O T E
Vials should not have a septum i.e., they should be open to avoid transfer of carryover
material on the septum.