1260 Infinity Binary LC - System User Guide
89
Quick Start Guide
5
Data Acquisition in Method and Run Control View
Table 9
Method parameters for first separation run - standard delay volume configuration
Module Parameter
Setting
Pump
Solvent A
Water
Solvent B
Acetonitrile
Flow rate
4.0 mL/min
Solvent
Composition
40 % A, 60 % B
Stop Time
1 min (1 min post time)
Autosampler
Injection
1 µL
Column
Compartment
Column
•
Column Eclipse Plus C18, 4.6 x 100 mm, 3.5 µm
(959961-902)
•
Column Poroshell 120 EC-C18, 3.0 x 50 mm, 2.7 µm
(699975-302)
•
Column Poroshell 120 EC-C18, 4.6 x 50 mm, 2.7 µm
(699975-902)
Temperature
40 °C
Detector
Signal A
Wavelength: 250 nm, Bandwidth: 4 nm
Reference Wavelength: 360 nm, Reference
Bandwidth: 100 nm