1260 Infinity Binary LC - System User Guide
137
Appendix
6
Setting Up a Method using Edit Entire Method
The access points in the work flow of the method are:
• Pre-Run Command / Macro
• Customized Data Analysis Macro
• Post-Run Command / Macro
Save Method with Data
saves a copy of the method in the data file and
names it RUN.M. This is not needed if the ChemStation is operated in the
usual configuration as the software always saves the method in the data
file (all versions since B.02.01). Only if the ChemStation has been
configured such that
Unique Sequence Folder Creation
is turned off and hence
methods are not routinely copied into the data file, this option will be
relevant.
As this is the final screen in the process, clicking
OK
will exit from
Run
Time Checklist
and from the
Edit Entire Method
process. The method should
now be saved into the master methods directory, by default
C:\Chem32\1\
Methods
, using
File > Save As > Method
or
Method Menu > Save Method As
.