88
1260 Infinity Binary LC - System User Guide
5
Quick Start Guide
Data Acquisition in Method and Run Control View
Data Acquisition in Method and Run Control View
Method Parameters for System Checkout with Isocratic Checkout
Sample
For checkout of the 1260 Infinity Binary LC, a test run with an isocratic
test mixture (Agilent isocratic checkout sample (01080- 68704)) will be
performed on the column ordered and delivered with your system.
Agilent isocratic checkout sample (01080- 68704) contains the following
components dissolved in methanol:
•
dimethyl phthalate
•
diethyl phthalate
•
biphenyl
•
o- terphenyl
The method parameters for separation of this test mixture are summarized
in
on page 89 for the standard delay volume configuration and in
on page 90 for the low delay volume configuration.