1260 Infinity Binary LC - System User Guide
3
Optimization of the Agilent 1260 Infinity Binary LC
How to Achieve Higher Resolution
Data Rate
The data rate setting for the UV detector must be selected appropriately.
Selecting a data rate that is too low results in increased peak width and
lower resolution, see
Figure 19
Peak width vs, UV data Rate
on page 59 shows the dependency of peak width, resolution and
peak capacity on the data rate. For fast and ultra- fast LC, high data rates
are needed; otherwise, the separation that is achieved on the column is
destroyed in the detector.