1260 Infinity Binary LC - System User Guide
25
The Agilent 1260 Infinity Binary LC - Product Description
1
Specifications
Composition
precision
< 0.15 % RSD or < 0.04 min SD whatever
is greater
at 0.2 and 1 mL/min; based
on retention time at constant
room temperature
Composition
accuracy
± 0.35 % absolute, at 2 mL/min, at
10 MPa (100 bar)
(water/caffeine tracer)
Control
Agilent control software (e.g.
ChemStation, EZChrom, OL, MassHunter)
Local control
Agilent Instant Pilot
Revision B.02.00 or above
Analog output
For pressure monitoring, 1.33 mV/bar,
one output
Communications
Controller-area network (CAN), RS-232C,
APG Remote: ready, start, stop and
shut-down signals, LAN optional
Safety and
maintenance
Extensive support for troubleshooting and
maintenance is provided by the Instant
Pilot, Agilent Lab Advisor, and the
Chromatography Data System.
Safety-related features are leak detection,
safe leak handling, leak output signal for
shutdown of pumping system, and low
voltages in major maintenance areas.
GLP features
Early maintenance feedback (EMF) for
continuous tracking of instrument usage
in terms of seal wear and volume of
pumped mobile phase with pre-defined
and user settable limits and feedback
messages. Electronic records of
maintenance and errors
Housing
All materials are recyclable
Table 2
Performance Specifications of the Agilent 1260 Infinity Binary Pump (G1312B)
Type
Specification
Comments
N O T E
For use with flow rates below 500 µl/min or for use without damper and mixer a vacuum
degasser is required.
All specification measurements are done with degassed solvents.