Optimization of the Agilent 1260 Infinity Binary LC
3
How to Achieve Higher Sensitivity
Choosing a Flow Cell
Several flavors of the Max- Light Cartridge Flow Cell are available, see
(
Normal Applications
The Max- Light Cartridge Cell ( 10 mm, V(s) 1.0 µL) (G4212- 60008) covers
a wide range of applications:
•
all column diameter down to at least 2.1 mm ID or even less
•
applications with peak dispersion (Peakwidth x flow) down to
~
2 µL
[example: pw = 0.04 min at flow = 0.1 mL/min gives peak dispersion of
0.04 min x 0.1 mL/min = 0.004 mL = 4 µL]
High Sensitiviy
If higher sensitivity is necessary, the Max- Light Cartridge Cell ( 60 mm,
V(s) 4.0 µL) (G4212- 60007) can be used. This cell enhances the detector
by lowering the limit of detection (LOD) by a factor of about 3 (depending
on the application).
Table 7
Specifications for Max-Light Cartridge Flow Cells
Cartridge Cells
•
Max-Light Cartridge Cell ( 10 mm, V(
s
) 1.0 µL) (G4212-60008)
•
Max-Light Cartridge Cell Bio-inert ( 10 mm, V(
s
) 1.0 µL) (G5615-60018)
•
Max-Light Cartridge Cell ( 60 mm, V(
s
) 4.0 µL) (G4212-60007)
•
Max-Light Cartridge Cell Bio-inert ( 60 mm, V(
s
) 4.0 µL) (G5615-60017)
•
HDR Max-Light Cartridge Cell ( 3.7 mm, V(
s
) 0.4 µL) (G4212-60032)
•
ULD Max-Light Cartridge Cell ( 10 mm, V(
s
) 0.6 µL) (G4212-60038)
•
Max-Light Cartridge Test Cell (G4212-60011)
Maximum
pressure
70 bar (1015 psi) Maximum Operating Pressure (MOP)
1
150 bar (2175 psi) Maximum Incidential Pressure (MIP)
2
1
Maximum Operating Pressure (MOP): The maximum pressure at which the system can operate
continuously under normal conditions.
2
Maximum Incidential Pressure (MIP): The maximum pressure which the system can experience
during a short time.
pH range
1.0-12.5 (solvent dependent)