80
Chapter 3, Programming the PCS Interface using GPIB Control
Set Up and Measure
Initialize
1.
Use the following command string to initialize the PCS Interface and query for
completion.
To PCS Interface>
SYST:PRES; *OPC?
NOTE:
The SYST:PRES command string will reset the PCS Interface but will maintain the current
temperature compensation. Do not use the *RST command during this initialization since
it will reset the temperature compensation. Compensations will be done at the first time
requested following the
*RST
command.
Set Up the UUT
ports
The internal signal path switches of the PCS Interface can be controlled to route
the Test Set’s connections to or from the UUT.
1.
Use the following command string for RX testing at the PCS Interface’s RF OUT only
port and for TX testing at the PCS Interface’s RF IN/OUT port.
To the PCS Interface>
RF:PATH 1
2.
Use the following command string for both RX and TX testing at the PCS Interface’s
RF IN/OUT port.
To the PCS Interface>
RF:PATH 2
NOTE:
When the RF:PATH command is executed, the TX and RX attenuators are automatically
set to their maximum values of 40 dB and 70 dB, respectively.
Содержание 83236B
Страница 7: ...vii...
Страница 10: ...x...
Страница 14: ...Contents 14 Error Message 154 Glossary 161...
Страница 30: ...30 Chapter 1 General Information Other Information...
Страница 123: ...123 5 S pecifications Specifications...
Страница 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...
Страница 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...
Страница 141: ...141 B M anual Changes Appendix B Manual Changes...
Страница 164: ...Index 164 TX definition 162 TX subsystem 114 TX RX port 25 typical 124 U upconverter definition 162 W warm up time 124...