154
Appendix C, Error Messages
Error Message
Error Message
0
(No error)
If an error was not generated and an error code was read at the GPIB, ’0’ is returned.
100
Value out of range; Rounding occurred
A numeric data parameter exceeded the range. Its value was rounded to a value within the
range.
101
Index out of range; Command ignored
The parameters of the commands were out of range.
102
Current path is invalid for the command
The command that was sent could not be executed due to the current
RF:PATH
specification.
−
100
Command error
A general syntax error that could not be classified more specifically was generated. This
occurs when a command error (defined in IEEE 488.2, 11.5.1.1.4) is generated.
−
101
Invalid character
A character that may not be used was included.
−
102
Syntax error
A command or data type that could not be recognized was used.
−
103
Invalid separator
A separator within a message (such as
;
or
,
) was invalid.