35
Chapter 2, Installation
When Used with Agilent 8924C and Controlled via Serial Port for CDMA Subscriber Unit
Tests
Cable Connections
Follow these steps to prepare the test equipment for use:
1
Set the GPIB Address Selector of the rear panel of the PCS Interface as follows.
GPIB/Ser switch
Ser
GPIB address
Any
2
Make the front and rear panel connections shown in figure 5.
3
Turn on the PCS Interface.
NOTE:
The PCS Interface must be powered on and connected before the Agilent 8924C is powered
on. If the PCS Interface is not powered on first, the Agilent 8924C will not detect it.
4
Set up the Test Set.
a
Turn on the Test Set.
b
Press and release the SHIFT key then the TESTS key to access the CONFIGURE
screen.
c
Position the cursor in the “PCS Intrfc Control” field, and press the knob to select
“On.”
d
Turn the Test Set’s power off and then back on.
5
Perform the RF Generator Level Calibration.
a
Press and release the SHIFT key then the TESTS key to access the CONFIGURE
screen.
b
Position the cursor in the “RF Gen Level” field, and press the knob to start the
calibration.
Содержание 83236B
Страница 7: ...vii...
Страница 10: ...x...
Страница 14: ...Contents 14 Error Message 154 Glossary 161...
Страница 30: ...30 Chapter 1 General Information Other Information...
Страница 123: ...123 5 S pecifications Specifications...
Страница 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...
Страница 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...
Страница 141: ...141 B M anual Changes Appendix B Manual Changes...
Страница 164: ...Index 164 TX definition 162 TX subsystem 114 TX RX port 25 typical 124 U upconverter definition 162 W warm up time 124...