110
Chapter 4, GPIB Commands
RX Subsystem
RX[:RFG1]:OUTPut:ATTenuator:MODE
ëû
{AUTO|HOLD}
Specifies the selection mode of the RX attenuator.
AUTO:
Auto mode.
(Default setting.)
HOLD:
Hold mode. The attenuator value can be set by the
RX:OUTP:ATT
command.
NOTE:
This command is not generally needed because the attenuator level is automatically
set when using the
RX:TSET:LEV?
or
RX:OUTP:LEV
command.
•
Response to the Query
{AUTO|HOLD} <new line><^END>
RX[:RFG1]:OUTPut:FREQuency
ëû
<
numeric
>[HZ|KHZ|MAHZ|MHZ|GHZ]
Specifies the frequency of the signal which you want to output from the
RF IN/OUT or RF OUT only port.
<numeric>:
800 to 960 MHz, 1710 to 1785 MHz, 1805 to 1910 MHz, or
1930 to 1990 MHz.
If the entered value is out of range, the value is
rounded off to the appropriate limit value. (The default unit is “Hz”.)
•
Response to the Query
{numeric (Hz)} <new line><^END>
NOTE:
Use the
RX:TSET:FREQ?
command which has the functions of both the
RX:OUTP:FREQ
and
RX:INP:FREQ?
commands.
Содержание 83236B
Страница 7: ...vii...
Страница 10: ...x...
Страница 14: ...Contents 14 Error Message 154 Glossary 161...
Страница 30: ...30 Chapter 1 General Information Other Information...
Страница 123: ...123 5 S pecifications Specifications...
Страница 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...
Страница 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...
Страница 141: ...141 B M anual Changes Appendix B Manual Changes...
Страница 164: ...Index 164 TX definition 162 TX subsystem 114 TX RX port 25 typical 124 U upconverter definition 162 W warm up time 124...