116
Chapter 4, GPIB Commands
TX Subsystem
TX:INPut:POWer:EXEC
Performs the temperature compensation for the power detector, measures the
average power of the burst of the signal input to the RF IN/OUT port, and stores
the value into the internal buffer memory. The trigger source which decides the
timing of the actual power measurement is set by the
TX:INP:POW:TRIG
command.
This command is valid only when the
RF:PATH
setting is 1 or 2.
TX:INPut:POWer:EXEC?
Returns the average power value measured by the
TX:INP:POW:EXEC
command
and stored into the internal buffer memory. (Query only)
•
Response to the Query
{numeric (dBm or W)} <new line><^END>
The unit suffix (dBm or W) is specified by the
TX:INP:POW:UNIT
command. (The
default is dBm.)
If the measured data is invalid,
−
130 (dBm) or 0 (W) is returned.
TX:INPut:POWer:PDETector:SAMPle:AVERage[:FACTor]
ëû
<
numeric
>
Specifies the number of bursts to be measured to calculate the average value of
the power measurement.
<numeric>:
1 to 10
If the entered value is out of range, the value is rounded off to the
appropriate limit value.
•
Response to the Query
{numeric} <new line><^END>
NOTE:
The number of samples to be collected is calculated by the following equation. The
LENGTH value is set by the
TX:INP:POW:PDET:SAMP:LENG
command and the
IGNORE value set by the
TX:INP:POW:PDET:SAMP:IGN
command.
Number of samples = LENGTH value * AVERAGE value
Содержание 83236B
Страница 7: ...vii...
Страница 10: ...x...
Страница 14: ...Contents 14 Error Message 154 Glossary 161...
Страница 30: ...30 Chapter 1 General Information Other Information...
Страница 123: ...123 5 S pecifications Specifications...
Страница 137: ...137 A F requency Conversion Tables Appendix A Frequency Conversion Tables...
Страница 140: ...140 Appendix A Frequency Conversion Tables Frequency Conversion for Signal Generation...
Страница 141: ...141 B M anual Changes Appendix B Manual Changes...
Страница 164: ...Index 164 TX definition 162 TX subsystem 114 TX RX port 25 typical 124 U upconverter definition 162 W warm up time 124...