Note:
Connect bare copper wires (approximately 3 cm in length)
to the
I
terminals of Channels 21 and 22 as shown below. These wires
will be used to make shorts across the channels in Tests 6 through 39.
Probe here for F501 measurement.
Probe here for F502 measurement.
Probe here
for L402
measurement.
Connections for 34901A Verification Tests 1 through 5
Note:
Be sure to probe the components
at the indicated locations on the module.
Chapter 4 Calibration Procedures
Relay Verification
90