Power Supplies (continued)
The A1 power supplies are tabulated below.
Power Supply
Minimum
Maximum
Switched
+5 Earth Ref.
4.75 V
5.25 V
No
+5 Backplane and Fan
4.75 V
5.25 V
Yes
+5 Floating
4.75 V
5.25 V
No
+18 Floating
17.6 V
19.9 V
No
-18 Floating
-19.0 V
-16.8 V
No
6 Vrms Filament
Yes
•
Check that the input to the supply voltage regulator is at least
1 V greater than its output.
•
Circuit failures can cause heavy supply loads which may pull down
the regulator output voltage.
•
Always check that the power supplies are free of ac oscillations
using an oscilloscope.
•
All plug-in modules use the +5 V backplane supply. Some plug-in
modules generate their own local power supplies.
Chapter 6 Service
Troubleshooting Hints
166
Содержание 34970A
Страница 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Страница 34: ...32 ...
Страница 35: ...2 Quick Start 2 ...
Страница 49: ...3 Front Panel Overview 3 ...
Страница 62: ...60 ...
Страница 63: ...4 Calibration Procedures 4 ...
Страница 118: ...116 ...
Страница 119: ...5 Theory of Operation 5 ...
Страница 159: ...6 Service 6 ...
Страница 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Страница 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Страница 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Страница 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Страница 182: ...180 ...
Страница 183: ...7 Replaceable Parts 7 ...
Страница 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Страница 221: ...8 Schematics 8 ...
Страница 272: ......