(Optional)
34903A Relay Contact Resistance Verification
1
Be sure to read “Plug-in Module Test Considerations” on page 87.
2
Install the 34903A module in slot 100. Close Channels 01 through 20.
Remove the module from the 34970A.
3
Measure the resistance from the
CM
terminal to the
NO
terminal on
each channel.
4
Install the module in slot 100. Open Channel 01 through 20.
Remove the module from the 34970A.
5
Measure the resistance from the
CM
terminal to the
NC
terminal on
each channel.
Note:
In general, a new relay should have a contact resistance of less than
0.2
Ω
. Relays with contact resistance in excess of 1.2
Ω
should be replaced.
Module Reference
4
Chapter 4 Calibration Procedures
Relay Verification
101
Содержание 34970A
Страница 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Страница 34: ...32 ...
Страница 35: ...2 Quick Start 2 ...
Страница 49: ...3 Front Panel Overview 3 ...
Страница 62: ...60 ...
Страница 63: ...4 Calibration Procedures 4 ...
Страница 118: ...116 ...
Страница 119: ...5 Theory of Operation 5 ...
Страница 159: ...6 Service 6 ...
Страница 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Страница 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Страница 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Страница 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Страница 182: ...180 ...
Страница 183: ...7 Replaceable Parts 7 ...
Страница 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Страница 221: ...8 Schematics 8 ...
Страница 272: ......