In addition to the +5 Volt power supply (Vcc) and ground, the module
controller uses four lines for control and communication:
•
RESET, from the Earth Referenced Logic A1U305. RESET is
common to all three slots. The module controller performs a reset
when this line goes high. Reset conditions vary for each plug-in.
•
SRQ, to the Earth Referenced Logic A1U305. The SRQ line is a
wired-OR line that can be driven by any plug-in. Consequently,
any module that asserts SRQ (line low), asserts this line in all
other slots and at the Earth Reference Logic.
•
DATA IN, from the Floating Logic A1U205 via the opto isolator
A1U312. This line is connected in common to all three slots.
•
DATA OUT, from the module controller to the Floating Logic
A1U205 via the opto isolator A1U213. This line is a wired-OR line
that can be driven by any module.
The DATA IN and DATA OUT lines are optically isolated from the
floating logic controller. These lines communicate with the Floating
Logic using an asynchronous serial bit stream.
The serial communications use an 11 bit protocol; a start bit, 8 data
bits, an attention bit, and a stop bit. The attention bit is 1 if the
8 data bits are an address/command, or 0 if the 8 data bits modify
or provide data for the previously sent command.
The module controller uses the hardwired slot-ID bits to decode the
serial bit stream address. When the address/command message
address matches the slot-ID, the plug-in is selected and responds to
the following commands. All other plug-ins will ignore the commands
until a new address/command message is received.
A 12 MHz crystal, Y101 is the clock for the module controller.
The module ID is a four bit pattern set through R106, R107, R108,
and R109. The Ferro-electric RAM U150 provides data storage of the
relay usage on the module (
see page 55)
and the thermocouple reference
junction temperature corrections (
see page
112). Data in and out of
U150 is serial.
The 34901A, 34902A, and 34908A each have two onboard, solid state
temperature sensors, U151 and U152, physically located near the
isothermal block at the input connections. The temperature sensors
are used as the thermocouple temperature reference.
Module Reference
5
Chapter 5 Theory of Operation
Switch Modules
139
Содержание 34970A
Страница 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Страница 34: ...32 ...
Страница 35: ...2 Quick Start 2 ...
Страница 49: ...3 Front Panel Overview 3 ...
Страница 62: ...60 ...
Страница 63: ...4 Calibration Procedures 4 ...
Страница 118: ...116 ...
Страница 119: ...5 Theory of Operation 5 ...
Страница 159: ...6 Service 6 ...
Страница 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Страница 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Страница 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Страница 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Страница 182: ...180 ...
Страница 183: ...7 Replaceable Parts 7 ...
Страница 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Страница 221: ...8 Schematics 8 ...
Страница 272: ......