Input Connections
You will need an input multiplexer module to verify or adjust the
internal DMM. Input connections can be made using a 34901A
20-Channel Multiplexer.
To use a 34901A to completely verify and adjust the internal DMM,
make the following connections:
Note:
Use shielded twisted pair PTFE
â
insulated cables to reduce settling
and noise errors. Connect the shield to the source LO output.
You can also use a 34902A for test and adjustment of voltage,
frequency, and resistance functions. You cannot test or adjust current
inputs with a 34902A. If you use a 34902A; connect the copper shorts
to Channels 7 and 15 and make the input connections to
Channels 8 and 16.
34901A
Connections for
DC/AC Current
Connections for DC V, AC V,
2- and 4-wire Ohms
Connections for
4-wire Ohms
To
Calibrator
Copper Short
Chapter 4 Calibration Procedures
Input Connections
66
Содержание 34970A
Страница 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Страница 34: ...32 ...
Страница 35: ...2 Quick Start 2 ...
Страница 49: ...3 Front Panel Overview 3 ...
Страница 62: ...60 ...
Страница 63: ...4 Calibration Procedures 4 ...
Страница 118: ...116 ...
Страница 119: ...5 Theory of Operation 5 ...
Страница 159: ...6 Service 6 ...
Страница 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Страница 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Страница 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Страница 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Страница 182: ...180 ...
Страница 183: ...7 Replaceable Parts 7 ...
Страница 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Страница 221: ...8 Schematics 8 ...
Страница 272: ......