–10 Vdc Adjustment Procedure (Optional)
The –10 Vdc calibration electronically enhances the Internal DMM’s
a-to-d converter linearity characteristic.
This adjustment should
ONLY be performed after servicing the A-to-D converter or
replacement of the calibration RAM.
You will need a 34901A 20-Channel Multiplexer to perform the
following procedures (
see page
66). Install the Multiplexer in slot 200.
1
If a zero calibration has not been performed recently, perform one
before beginning this procedure (
see page 81
).
2
Select Channel 210. Configure the channel as follows:
DC VOLTS
10 V range
6
1
⁄
2
digits
INTEG 100 PLC (in the
Advanced
menu)
INPUT R > 10 G (in the
Advanced
menu)
Before executing each test, you must press
to enable reading
monitoring on the selected channel (or use the
ROUTe:MON
command
from the remote interface).
3
Measure and note the voltage offset present at the end of the
measurement cable by shorting the ends of the Channel 210
measurement cable. Be sure to use a copper wire and allow enough
time for the residual thermal offset to stabilize (usually about
1 minute).
4
Connect the input cable to the calibrator output and set the calibrator
to 10V. Allow enough settling time for any thermal offset
voltages to stabilize (usually about 1 minute).
5
Perform a +10V dc gain calibration. Press
to enter the
calibration menu. Press
again to begin the adjustment procedure.
6
Use the knob,
and
to set the number in the display to the
sum of the calibrator output and the measured offset (from step 3)
and press
. For example, if the calibrator output is 10.001 volts
and the measured offset is 10
µ
V, enter +10.001010 volts. When the
adjustment finishes, verify that new readings fall within
±
20
µ
V of
the calibrator output plus the offset.
Continued on next page...
4
Chapter 4 Calibration Procedures
–10 Vdc Adjustment Procedure (Optional)
85
Содержание 34970A
Страница 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Страница 34: ...32 ...
Страница 35: ...2 Quick Start 2 ...
Страница 49: ...3 Front Panel Overview 3 ...
Страница 62: ...60 ...
Страница 63: ...4 Calibration Procedures 4 ...
Страница 118: ...116 ...
Страница 119: ...5 Theory of Operation 5 ...
Страница 159: ...6 Service 6 ...
Страница 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Страница 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Страница 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Страница 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Страница 182: ...180 ...
Страница 183: ...7 Replaceable Parts 7 ...
Страница 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Страница 221: ...8 Schematics 8 ...
Страница 272: ......