(Optional)
34905A/06A Relay Contact Resistance Verification
Note:
Be sure to use the correct SMB connectors (50
Ω
or 75
Ω
).
1
Be sure to read “Plug-in Module Test Considerations” on page 87.
2
Prepare the module by connecting an SMB short to
CH10
,
CH11
,
CH12
, and
CH13
. Connect the
COM1
terminal to the DMM.
Be sure to use the correct SMB connectors for the module.
3
Install the module in slot 100.
4
Close Channel 11.
5
Measure the resistance on the DMM.
6
Repeat steps 4 and 5 for Channels 12, 13, and 14.
7
Repeat steps 2, 3, 4, 5, and 6 for the channels connected to
COM2
.
Note:
In general, a new relay should have a contact resistance of less
than 0.5
Ω
. Relays with contact resistance in excess of 1
Ω
should be replaced.
Module Reference
4
Chapter 4 Calibration Procedures
Relay Verification
105
Содержание 34970A
Страница 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Страница 34: ...32 ...
Страница 35: ...2 Quick Start 2 ...
Страница 49: ...3 Front Panel Overview 3 ...
Страница 62: ...60 ...
Страница 63: ...4 Calibration Procedures 4 ...
Страница 118: ...116 ...
Страница 119: ...5 Theory of Operation 5 ...
Страница 159: ...6 Service 6 ...
Страница 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Страница 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Страница 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Страница 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Страница 182: ...180 ...
Страница 183: ...7 Replaceable Parts 7 ...
Страница 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Страница 221: ...8 Schematics 8 ...
Страница 272: ......