Plug-in Module Test Considerations
For optimum performance, all test procedures should comply with the
following recommendations:
•
Assure that the calibration ambient temperature is stable and
between 18
°
C and 28
°
C. Ideally the calibration should be
performed at 23
°
C
±
1
°
C.
•
Assure ambient relative humidity is less than 80%.
•
Install the plug-in module and allow a 45 minute warm-up period
before verification or adjustment.
•
Use shielded twisted pair PTFE insulated cables to reduce settling
and noise errors. Keep the input cables as short as possible.
•
Remove all user wiring and connections from the plug-in modules
before verification or adjustment.
•
Use 4-wire Ohms measurement techniques for checking relay
contact resistance. Check directly at the terminals where possible.
Module Reference
4
Chapter 4 Calibration Procedures
Plug-in Module Test Considerations
87
Содержание 34970A
Страница 1: ...Agilent 34970A Data Acquisition Switch Unit Service Guide ...
Страница 34: ...32 ...
Страница 35: ...2 Quick Start 2 ...
Страница 49: ...3 Front Panel Overview 3 ...
Страница 62: ...60 ...
Страница 63: ...4 Calibration Procedures 4 ...
Страница 118: ...116 ...
Страница 119: ...5 Theory of Operation 5 ...
Страница 159: ...6 Service 6 ...
Страница 177: ...General Disassembly 4 3 2 1 6 Chapter 6 Service Disassembly 175 ...
Страница 178: ...Internal DMM Disassembly 3 2 1 Chapter 6 Service Disassembly 176 ...
Страница 180: ...Additional Chassis Disassembly 1 2 Chapter 6 Service Disassembly 178 ...
Страница 181: ...Plug In Module Disassembly Needle nose Pliers 6 Chapter 6 Service Disassembly 179 ...
Страница 182: ...180 ...
Страница 183: ...7 Replaceable Parts 7 ...
Страница 220: ...Chapter 7 Replaceable Parts Manufacturer s List 218 ...
Страница 221: ...8 Schematics 8 ...
Страница 272: ......