74
34410A/11A User’s Guide
2
Features and Functions
Automatic Trigger Delay
If you do not specify a trigger delay manually, the multimeter selects the
trigger delay time automatically. Automatic trigger delays are set to ensure
that completely settled and accurate measurements are taken. The delay is
determined by function, range, integration time and ac filter setting, as
shown in the following tables
DC Voltage Measurements (all ranges):
DC Current Measurements (all ranges):
2
–
wire Resistance Measurements :
4
–
wire Resistance Measurements
Integration Time
Trigger Delay
¡
1 PLC
160
m
s
0.2 PLC, 0.06 PLC
130
m
s
~
0.02 PLC
100
m
s
Integration Time
Trigger Delay
¡
1 PLC
1.5 ms
~
0.2 PLC
1.0 ms
Range
Trigger Delay
(for
~
0.02 PLC)
Trigger Delay
(for 0.06 or 0.2 PLC)
Trigger Delay
(for
¡
1 PLC)
100
W
80
m
s
100
m
s
130
m
s
1
k
W
110
m
s
130
m
s
160
m
s
10 k
W
130
m
s
160
m
s
190
m
s
100 k
W
540
m
s
670
m
s
800
m
s
1 M
W
5.0 ms
6.0 ms
7.5 ms
10 M
W
–
1G
W
60 ms
70 ms
84 ms
Range
Trigger Delay
(for
~
0.02 PLC)
Trigger Delay
(for 0.06 or 0.2 PLC)
Trigger Delay
(for
¡
1 PLC)
100
W
–
100 k
W
1.0 ms
1.0 ms
1.5 ms
1
M
W
10 ms
10 ms
15 ms
10 M
W
–
1 G
W
100 ms
100 ms
100 ms
Содержание 34410A
Страница 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Страница 6: ...6 34410A 11A User s Guide...
Страница 12: ...12 34410A 11A User s Guide...
Страница 34: ...34 34410A 11A User s Guide 1 Quick Start...
Страница 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Страница 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 135: ...34410A 11A User s Guide 135 Specifications 5 Dimensions All dimensions are shown in millimeters...
Страница 140: ...140 34410A 11A User s Guide 5 Specifications...