34410A/11A User’s Guide
79
Features and Functions
2
Data Logging
The Data Logger function makes the multimeter a programmable tool for
taking and storing a series of timed measurement readings. This feature
can be used in combination with any primary measurement function. The
readings are stored in non–volatile memory, and can be accessed from the
front panel or remote interface.
To customize a data logging event, the multimeter allows you to set the
following parameters:
START DELAY
,
INTERVAL
,
TIME/EVENT
(
DURATION
or
COUNT
).
•
START DELAY
is the delay time, settable from 0 to 3600 seconds, from
initiating a trigger before the data logging function takes its first
reading.
•
INTERVAL
is the time, settable from 0 to 3600 seconds, between
subsequent readings.
•
TIME/EVENT
is the total duration of the data logging event. You may
choose to set a
DURATION
in hours, minutes and seconds, from 1 second
to 100 hours, or a
COUNT
of total readings, settable from 1 to 50,000
readings (model 34410A) or up to 1 million readings (model 34411A).
•
Front Panel Operation:
To set up a data logging event, press
to
display
DATA LOGGER
. Select
SETUP
and press
to display
START DELAY
.
Use the navigation keypad to set the desired start delay, and press
to display
INTERVAL
. Use the navigation keypad to set the desired reading
interval, and press
to display
TIME/EVENT
. Press
or
as
needed to select
DURATION
or
COUNT
, and press
.
•
If
DURATION
was selected, use the navigation keypad to set the desired
reading interval, and press
.
If
COUNT
was selected, use the
navigation keypad to set the desired total number of readings, and
press
.
•
Upon pressing
, the displays will briefly read
TO START
and
PRESS
TRIGGER KEY
. The data logger is now programmed. The primary display
line will then show the selected primary measurement function. The
second display line will show a progress bar and the count of set
total readings. These displays will remain static until you press
,
or pulse the rear panel Ext Trig terminal.
•
When you press
, the multimeter will begin to take readings,
after the set start delay has expired. While it is reading, the primary
display line will show each reading, and the second display line will
show a progress bar and a countdown of readings remaining. The
multimeter will continue taking readings until the programmed total
readings count has been reached, or unless the data logger is stopped
manually.
Содержание 34410A
Страница 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Страница 6: ...6 34410A 11A User s Guide...
Страница 12: ...12 34410A 11A User s Guide...
Страница 34: ...34 34410A 11A User s Guide 1 Quick Start...
Страница 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Страница 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 135: ...34410A 11A User s Guide 135 Specifications 5 Dimensions All dimensions are shown in millimeters...
Страница 140: ...140 34410A 11A User s Guide 5 Specifications...