78
34410A/11A User’s Guide
2
Features and Functions
Reading Memory
This feature is available from the remote interface only.
The multimeter’s reading memory is a first–in, first out (FIFO) buffer
holding up to 50,000 readings (model 34410A) or 1 million readings
(model 34411A). The oldest readings are preserved.
•
Remote Interface Operation:
The following command transfers readings
stored in non–volatile memory into the multimeter’s output buffer, from
where you can read them into your computer:
FETCh?
The following command will obtain whatever readings are present in
NVMEM, and erase them as they are read:
R?
Refer to the
Agilent 34410A/34411A Programmer’s Reference
for
complete description and syntax for these commands.
Содержание 34410A
Страница 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Страница 6: ...6 34410A 11A User s Guide...
Страница 12: ...12 34410A 11A User s Guide...
Страница 34: ...34 34410A 11A User s Guide 1 Quick Start...
Страница 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Страница 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 135: ...34410A 11A User s Guide 135 Specifications 5 Dimensions All dimensions are shown in millimeters...
Страница 140: ...140 34410A 11A User s Guide 5 Specifications...