34410A/11A User’s Guide
139
Specifications
5
Configuring for Highest Accuracy Measurements
The measurement configurations shown below assume that the multimeter
is in its power
–
on or reset state. It is also assumed that auto–ranging is
enabled to ensure proper full scale range selection.
DC Voltage, DC Current, and Resistance Measurements:
•
Select
NPLC
and
100
(NPLCs) for
INTEGRATION
.
•
Set
INPUT Z
to
HI
–
Z
(for the
100 mV
,
1 V
, and
10 V
ranges) for the best dc
voltage accuracy.
•
Use the 4
–
wire ohms function (
W
4W) for the best resistance
measurement accuracy.
•
For 2
–
wire ohms, dc voltage and dc current measurements, set
AUTOZERO
to
ON
to remove thermal EMF and offset errors.
•
Null the test lead resistance for 2
–
wire ohms measurements, and to
remove any interconnection offset for dc voltage measurements.
AC Voltage and AC Current Measurements:
•
Set the
AC FILTER
to
3 Hz: SLOW
.
Frequency and Period Measurements:
•
Set the
GATE TIME
to
1 sec
.
Содержание 34410A
Страница 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Страница 6: ...6 34410A 11A User s Guide...
Страница 12: ...12 34410A 11A User s Guide...
Страница 34: ...34 34410A 11A User s Guide 1 Quick Start...
Страница 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Страница 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 135: ...34410A 11A User s Guide 135 Specifications 5 Dimensions All dimensions are shown in millimeters...
Страница 140: ...140 34410A 11A User s Guide 5 Specifications...